TRANSPORT VENTILATOR EPV-200 ALLIED HEALTHCARE
The EPV200 Portable Ventilator with Assist Control provides ventilation for intubated and non-intubated patients, in emergency situations where medical surge response is needed. This unit is gas-powered and electronically-controlled, designed to be user-friendly for rapid deployment. Settings visible on the unit include inspiratory time, tidal volume, BPM controls, digital manometer, along with visual and audible safety alarms. The ventilator is also equipped with an Assist-Control that triggers at less than -2 cm H2O for regulating spontaneous breathing.
Product Information
Standard Configuration:
- Dimensions: 3.5″ x 7″ x 9.3″ (8.9 cm x 17.8 cm x 23.6 cm).
- Weight: 3.1 lb (1.4 kg) with batteries.
- Temperature:
- Operating: -15 to 122 °F (-9 to 50 °C).
- Storage: -40 to 140 °F (-40 to 60 °C).
- Enclosure: ABS plastic with seals prevent water ingress.
- Shock Resistance: 30″ drop.
- Vibration Resistance: 100 G.
- Water Resistance: Spill and rain resistant.
Power Characteristics:
- Power Source: Pneumatic with electronic controls and alarms.
- Power Input: Pneumatic. 40 – 87 psi O2 and 2 D cell batteries.
- Internal Battery Type: D Cell batteries.
- Operating Time: Based on average adult settings of 10 BPM, 640 ml tidal volume, 2 second inspiratory time.
- Oxygen: Dependent on source capacity. Approx. 65 minutes on D cylinder.
- Battery Life: 48 hours of continuous use.
Setting Parameters:
- Ventilation Mode: Assist-Control.
- Assist-Control: Triggers at less than -2 cm H2O.
- Flow Rate: 12 – 36 LPM.
- Breathing Rate: Adjustable. 0, 5 – 30 breaths per minute.
- Tidal Volume: Adjustable. 200 – 1200 ml.
- Inspiratory Time: Selectable. 1 or 2 seconds.
- PEEP: External. 0 – 20 cm H2O (with PEEP adapter).
- FIO2: 100%.
- Manometer Readout: Digital. 0 – 99 cm H2O.
- Airway Pressure Limit: Fixed. 60 cm H2O.
Alarms:
- High Airway Pressure: Fixed. 45 cm H2O.
- Low Airway Pressure / Circuit Disconnect: Fixed. 9 cm H2O.
- Low Source Gas: Fixed. Activates at approx. 40 psi (275 kPa).
- Low Battery: Activates when approx. 2 hours run time remain.
- Alarm Volume: Approx. 60 dB at 1 meter.
- Alarm Silence: Yes. 110 seconds.


ReviewsThere are no reviews yet.