BABY RESUSCITATION TROLLEY ECOSY-930 BNG MEDICAL
Utilize Elstein far infrared ceramic heating tube,
which greatly improve the radition efficiency and
extend the duration.
Perfect to absorbed by skin and effectively
accelerate blood circulation, by which the visceral
temperature and body surface temperature rise
simultaneously in short time.
Double heating tube design accelerate the heating
rate and ideally distribute the heat on the surface
even the mattress is tilted, which provides warm
and comfortable for neonatal.
Apply high-accurate double thermistor
temperature sensor and increase its accuracy of
temperature monitoring.
Product Information
Standard Configurations:
- Power requirements: Customer specified
- Power consumption: 700VA (Maximum)
- Heater capacity: 550W
- Dimensions: 84(W) X ll0(D) X 178-218(H)cm (Mattress surface: 80-120cm)
- Weight: 100kg
- Temperature control: Servo control (Manual control also available)
- Skin temperature: Setting range : 34.0-38.0°C
- Display range : 30.0-4 .0°C
- Heater output: Setting range: 0-100% (in 5% increments)
- Indication range : 0-100 (in 10 levels)
- Timer: Time display range : 0-60minutes 00 seconds (in 1 second increments) APGAR mode /CPR mode switchable
- Lighting: LED lamp Adjustable illumination intensity 0-1200lux
- Mattress: Platform angle Inclination angle (back and fo11:h): +13° ~ -13°
Optional:
Phototherapy Unit
X-Ray Cassette Tray
Hydraulic Up & Down
Weight Scale
- Weight range: 300g to 8kg
- Weight display resolution: 1 g
- Weight accuracy: +/-10 g
Pulse Oximeter Unit
- Display range: 1-100% SpO2
- SpO2 alarm setting range: Upper limit 0-99 % (i.11 1 % increment ) Lower limit : 4 -9 % (in 1 % increment )
- Pulse rate display range: 2 – 40bpm
- Pulse rate alarm setting range: Upper limit : 80- 40bpm, OFF Lower limit : 3 -1 S0bpm, OFF (in bpm increments)
- Alarms: SpO2 high/low limit, pulse rate high/low limit, SpO2 sensor
Resuscitation Unit
- Resusci-flow unit: PMAX setting range : 19 – 8 kPa ( 0 ~ 60cmH2O) Default setting range: 39 kPa (40cmH2O)
Manometer
- Manometer: Pressure indication: -2 ~ 8 kPa (-20 ~ 80cmH2O)
Suction Unit
- Suction unit: Suction pressure control range: 0 ~ 6. kPa (0 ~ 200nunHg)
Oxygen Air Blender
- O2 Blender: Oxygen concentration control range : -1 ~ 100% Oxygen flow rate control range : 0 ~ 1 L/min.







ReviewsThere are no reviews yet.