| SPO2 |
|
| Adult / Pediatric |
70 to 100% +/-2 digits SpO2 |
| Neonate |
70 to 100% +/-3 digitsSpO2 |
| Temperature |
|
| Scale |
Selectable C° or F° Range: 15° to 45° C/ 59° to 113° F |
| Accuracy |
+/- 0.1°C (15° C to 45° C); +/- 0.2° F (59° to 113° F) exclusive of probe errors |
| NIBP |
|
| Technique |
Oscillometric |
| Connector Type |
Rectus |
| Systolic Range |
adult: 55-235 mmHg, pediatric: 55-160 mmHg, neonate: 45-120 mmHg |
| Diastolic Range |
adult: 30-220 mmHg, pediatric: 30-150 mmHg, neonate: 20-100 mmHg |
| Systolic Accuracy |
mean error less than +/- 5 mmHg, standard deviation less than +/- 8 mmHg |
| Diastolic Accuracy |
mean error less than +/- 5 mmHg, standard deviation less than +/- 8 mmHg |
| Pulse Rate Range |
adult/pediatric: 35 to 245 bpm, neonate: 70 to 245 bpm |
| Pulse Rate Accuracy |
+/-3 bpm or 3% whichever is greater |
| Respiration |
|
| Range |
4 to 200 bpm |
| Accuracy |
+/-2% or 2 breaths per minute whichever is greater from 4 to 150 bpm, +/- 4% from 151 to 200 bpm |
| ECG |
|
| ECG Leads |
I, II, III, aVR, aVL, aVF, V |
| Respiration Accuracy |
+/- 1 breath per minute from 0-40 bpm, +/-2 bpm from 41-70 bpm, +/-3 bpm from 71-100 bpm, +/-5 bpm from 101-150 bpm |
| Respiration Rate |
0-150 bpm |
| Accuracy |
0-38 mmHg +/-2 mmHg, 39-99 mmHg +/-5% +0.08% for every 1 mmHg above 40 mmHg |
| Range |
0-99 mmHg |
| IBP |
|
| Range |
Sys/Dia/Mean: 30- +300 Mmhg |
| Accuracy |
+/- 2 Mmhg Or 2% Whichever Is Greater |
| Scale |
-10 To 10, 0-20, 0-40, 0-160, 0-225, 0-320, 60-140 Mmhg |
| Heart Rate Meter |
|
| Range |
30-300 Bpm Adult/ Pediatric, 30-350 Bpm Neonate |
| Accuracy |
+/-3bpm Or +/-3% At 30-250 Bpm Whichever Is Greater, +/-5% At 251-350 Bpm |
ReviewsThere are no reviews yet.